Menu
Halogen acid gas Release Measuring Device AH68016
Halogen acid gas Release Measuring Device AH68016 is used for testing the content of the released halogen acid gas, PH value (hydrogen ion content), conductivity etc. when the cable is burning.
Activated carbon (air filter) | Filter air source by the filter |
---|---|
Silica gel (air dry) | Device for drying air source |
Flow-meter | 0.1-1L/min adjustable |
Quartz tube | Φ40*800 |
Details About Halogen acid gas Release Measuring Device AH68016
Standard:
GB/T 17650.1-2-1998
IEC 60754-1-2-1994
Technical parameters:
- Activated carbon (air filter): Filter air source by the filter.
- Silica gel (air dry): Device for drying air source
- Flow-meter: 0.1-1L/min adjustable, device for controlling the flow of air source.
- Thermocouple: K type of high temperature resistant, stainless thermocouple, 0-1300℃
- Quartz tube: Φ40*800
- Magnetic stirrer: A stirrer used for stirring the burning gas in distilled water.
- PH meter: A device used for measuring the PH acidity (ray magnetic).
- Conductor meter: An instrument used for measuring the conductivity of the burning gas (ray magnetic).
- 7 “color touch screen and PLC and PID control heating furnace temperature automatic calculation module.
- Modes’ swapping can meet 2 kinds of testing methods.
- Heating furnace: Φ220*600, effective heating space: Φ43*550, power: 4KW.
- Washing bottle: 500ml, straight-pipe washing bottles, 2 pcs, measuring pot of 2L, 1 pcs..
Specification:
- Installation dimension: 2300*600 mm
- Power: AC 220V, 50/60Hz, 21A
- Reagent and gas source: Customer-owned.
Share Halogen acid gas Release Measuring Device AH68016 With Others
Share on facebook
Share to Facebook
Share on twitter
Share to Twitter
Share on linkedin
Share to LinkedIn
Share on pinterest
Share to Pinterest